PCE Instruments

3500 - Measuring instrument PCE Instruments - Free user manual and instructions

Find the device manual for free 3500 PCE Instruments in PDF.

📄 50 pages English EN Download 💬 AI Question 10 questions ⚙️ Specs
Notice PCE Instruments 3500 - page 2
Pick your language and provide your email: we'll send you a specifically translated version.
Product Type Hardness Tester (Portable)
Brand PCE Instruments
Model PCE-3500
Measurement Methods UCI (Ultrasonic Contact Impedance) and Rebound (Leeb)
Hardness Scales Rockwell C, Brinell, Vickers, Leeb, Tensile Strength (via Brinell), and custom user scales
Calibration Scales 88 scales across 8 types and 11 materials
Information Processing Unit Dimensions 160 x 75 x 30 mm
UCI Probe Dimensions ∅25 x 140 mm
Rebound Probe Dimensions ∅20 x 150 mm
Weight (Processing Unit) 0.25 kg
Weight (UCI Probe) 0.125 kg
Weight (Rebound Probe) 0.25 kg
Power Supply Three 1.5V AA rechargeable batteries (NiCd or NiMH)
Battery Life Approx. 6 hours (NiCd) or 10 hours (NiMH)
Operating Temperature -20°C to +40°C
Display LCD with adjustable brightness
Memory Up to 60,000 measurements
Data Output USB cable for PC transfer
Language Options English, Russian
Package Contents Processing unit, UCI probe (10N or 50N), rebound probe, charger, batteries, USB cable, manual, hardness test blocks
Maintenance Clean probe tip with soft cloth and alcohol; avoid water and solvents; store in carry case with batteries removed if idle >14 days
Safety Warnings Use only rechargeable batteries; avoid sharp impacts on diamond indenter; keep away from aggressive chemicals and direct sunlight; do not immerse in liquids
Spare Parts / Repairability Contact manufacturer or authorized service center; warranty covers defects within one year (excludes battery, charger, bag)

Frequently Asked Questions - 3500 PCE Instruments

What measurement methods does the PCE-3500 use?
The PCE-3500 uses two methods: UCI (Ultrasonic Contact Impedance) with a Vickers diamond indenter oscillating at ~78 kHz, and Rebound (Leeb) where an impact body is propelled against the surface and its rebound speed is measured.
What hardness scales are available?
The device supports Rockwell C, Brinell, Vickers, Leeb, Shore, and Tensile Strength (calculated from Brinell). It also includes 8 user-defined scales (User 1-3) and 11 material groups (steel, alloy steel, stainless steel, cast iron, nodular iron, aluminum, brass, bronze, copper, user materials 1-2).
What are the minimum specimen requirements for UCI measurement?
For UCI probe: surface roughness Ra ≤ 1.5 µm (10N) or 2.5 µm (50N), minimum radius of curvature 5 mm, weight ≥ 0.1 kg, thickness ≥ 1 mm. For thin or light specimens, fix them to a solid base with viscous paste to avoid resonance.
What are the minimum specimen requirements for Rebound measurement?
For Rebound probe: surface roughness Ra ≤ 3.2 µm, minimum radius of curvature 10 mm, weight ≥ 5 kg (if 2-5 kg, must be coupled to a large support; less than 2 kg must be firmly fixed). Minimum thickness 10 mm.
How do I perform a UCI measurement?
Select UCI mode, choose the scale and material. Place the probe puck flat on the sample, press the skirt to bring the diamond tip perpendicular to the surface, then slowly apply force (1 or 5 kg depending on probe) until a beep sounds. Avoid sharp clicks that could damage the diamond.
How do I perform a Rebound measurement?
Select Rebound mode, set the measurement angle. Place the probe vertically on the test surface, press down until the charging mechanism clicks, then press the release button. A beep indicates the measurement; the hardness value appears on the display.
How do I calibrate the device?
Use three certified hardness blocks covering low, medium, and high hardness. In the Calibration menu, select the cell for the desired scale and material. Enter the nominal values of the blocks, then take at least 5 measurements on each block. Save the calibration. Additional user calibration (1 or 2 points) can be applied later if needed.
What measurement modes are available?
The device offers Graph (real-time or average plot), Histogram (bar chart), Statistics (max, min, deviation, average, count), Smart (filters outliers based on a tolerance percentage), and Signal (shows voltage waveform for Rebound probe).
How do I maintain the probe and device?
Clean the probe tip (diamond or carbide ball) with a soft cloth moistened with alcohol. Do not use water or solvents on the processing unit; wipe with a dry cloth. Check probe operation regularly using reference blocks. Store the tester in its carry case, and remove batteries if storing for more than 14 days.
What should I do if the device gives erratic readings?
First, ensure the batteries are fully charged (low battery can cause errors). Check that the probe is clean and undamaged. Verify the material and scale settings match the sample. If the problem persists, recalibrate using certified test blocks. If still faulty, contact PCE Instruments service.

User questions about 3500 PCE Instruments

0 question about this device. Answer the ones you know or ask your own.

Ask a new question about this device

The email remains private: it is only used to notify you if someone responds to your question.

No questions yet. Be the first to ask one.

Download the instructions for your Measuring instrument in PDF format for free! Find your manual 3500 - PCE Instruments and take your electronic device back in hand. On this page are published all the documents necessary for the use of your device. 3500 by PCE Instruments.

USER MANUAL 3500 PCE Instruments

Southpoint Business Park

Ensign way

Hampshire / Southampton

United Kingdom, SO31 4RF

From outside UK: +44

Tel: (0) 2380 98703 0

Fax: (0) 2380 98703 9

info@pce-instruments.com

www.pce-instruments.com/english

www.pce-instruments.com

Bedienungsanleitung

Durometer

PCE-3500

PCE-3500 UCI Hardness Tester HRC 61.8 Graph Steel MODE MAT SCALE OK BACK BS ENTER

PCE Instruments 3500 - PCE-3500 - 2

natural_image Close-up of a metallic cylindrical device with textured surface and attached probe, mounted on a black base (no visible text or symbols)

CONTENTS

  1. Introduction 2
  2. Appointment 2
  3. Specifications 2
  4. Packing list 5
  5. Principle of operation 5
  6. Preparations for operation 7
  7. Basics of operation 8
  8. Operation 9
  9. Hardness measurement 19
  10. Measurement modes 30
  11. Calibration 35
  12. Technical maintenance, special conditions of operation 46
  13. Precautions and trouble shouting 48
  14. Manufacturer's guarantee and service maintenance 49

1. Introduction

The following operation manual explains the preparation, setup, principles of operation, usage, and troubleshooting of the hardness tester PCE-3500.

Please, read this instructions carefully for operate the hardness tester PCE-3500 functions quickly and effectively.

In doing this you will be able to take full advantage of the function range of the instrument. At the same time, you will also avoid errors and wrong operation which in turn would cause incorrect test results and thus could lead to injury and damage.

2. Appointment

The hardness tester PCE-3500 is handy, easy to operate and can carry out tests quickly without any difficulties.

Hardness tester PCE-3500 is mainly suitable:

• for measurement hardness of low and non-alloyed steels;
• for measurement hardness of high-alloyed steels;
• for measurement hardness of nonferrous metals.

3. Specifications

3.1. Calibration scales

Device has 88 calibration scales of hardness that conventionally divided into eight scales and eleven materials for each:

07:31 HRC HB HV HRB HS U1 U2 U3 ST AST SST CI NI AL BRS BRZ CU U1 U2 Empty cell Calibrated User calibration

Each of the scales can be additionally calibrated by 1 or 2 points. Also, the unit measures in Leeb scale (with rebound probe) and Tensile strength, through the automatic recalculation of the Brinell scale.

3.2. The limits of permissible basic error of measurement

Hardness scaleError
Rockwell C±2
Brinell±10
Vickers±15
Leeb±15
Tensile strengthNot rated

3.3. Working conditions: from -20°C to +40°C

3.4. Overall dimensions

NameOverall dimensions, mm
Information processing unit160x75x30
UCI probe∅25x140
Rebound probe∅20x150

3.5. Weight of the instrument and probes

NameWeight, kg
Information processing unit0,25
UCI probe0,125
Rebound probe0,25

3.6. Power supply

Battery operation: three 1,5V AA rechargeable batteries.

NiCd (approx. 6 hours operation) or NiMH (approx. 10 hours operation).

NOTE: Do not allow using of non-rechargeable cells!

3.7. To save battery charge in the menu there are settings of device auto switch off and brightness of the display.

3.8. Requirements of the test material

Surface roughness, not more, Ra
UCI (10N) probe U11,5
UCI (50N) probe U12,5
Rebound probe D13,2
Radius of curvature of the surface, mm
UCI probe U15
Rebound probe D110
Weight of the test material, not less, kg
UCI probe U10,1
Rebound probe D15
Thickness of the test material, not less, mm
UCI probe U11
Rebound probe D110

4. Packing list

4.1. Information processing unit 1 pc
4.2. Probes:

UCI (10N) ____ pc

UCI (50N) _ pc

Rebound (Leeb) ____ pc

4.3. Charger 1 pc
4.4. Battery 3 pc
4.5. USB cable 1 pc
4.6. Operation manual 1 pc
4.7. Hardness test block __pc

5. Principle of operation

5.1. Rebound method

Rebound probe consists of an impact body and the capture. The impact body has a carbide tip and a permanent magnet for generating a voltage pulse; the impact device has a spring mechanism for loading and impelling the impact body, and an induction coil for detecting the magnet in the impact body. In the rebound hardness testing method, the speed variation caused by the impact of the impact body against the material surface is measured.

The impact energy is adjusted via the spring for the measurement. The impact body contained in the tube of the impact device is impelled against the test surface by means of the release button. In the course of this, the magnet of the impact body induces in the coil a voltage signal whose height is proportional to the impact phase speed. The impact causes a plastic deformation of the material and a permanent spherical indentation is produced in the surface. This

plastic deformation leads to a loss of energy of the impact body and thus to a lower speed after the actual rebound phase.

The speed ratio is determined exactly at the moment of impact/rebound by means of the special signal processing. The speed ratio is therefore unaffected by the impact direction. As opposed to this, other rebound hardness testers require presetting of the impact direction in fixed steps (influence of gravitation on the speed ratio) - which constitutes a considerable disadvantage with frequently changing test positions.

5.2. UCI method

The Vickers diamond is fixed to the tip of a round metal rod. This metal rod is excited, to its resonant frequency of approx. 78 kHz, into longitudinal oscillations. When the Vickers diamond contacts the sample surface, the resonant frequency will change. This change happens in relation to the size of the indent area from the Vickers diamond. The size, in turn, is a measure for the hardness of the tested material. Resonant frequencies can be measured very accurately. This is why the UCI method is suited to make the evaluation of Vickers indents, and thus of the complete test procedure, so much easier and quicker.

There are also two additional advantages:

- the measurement is made under load. (No impairment of the measurement due to elastic resilience);

- the hardness measurement is based on the area of the indent and not on the length of the indent diagonals.

The measurement is thus less affected by surface roughness; even gunmetal-finished surfaces can be measured.

Concerning the UCI method, the measurement value is also dependent on the Young's modulus of the material.

6. Preparations for operation

6.1. Battery supply

The hardness tester PCE-3500 is powered by accumulators. You will need three AA 1,5V rechargeable (Nickel-Cadmium or Nickel-Metalhydrid) batteries.

NOTE! The batteries must be fully charged before first use.

Use of the batteries.

  • Open the battery compartment.
  • Insert the batteries, observing the correct polarity.
  • Close the battery compartment.

Used or defective batteries are special refuse and must be disposed of according to the governing laws!

6.2. Connection of probes.

Socket of the probe is at the top of the information processing unit. Connect the probe cable to the PCE-3500 socket in accordance with the marks on the connector.

7. Basics of operation

7.1. Keys
Turn on/off key PCE-3500 UCI Hardness Tester Materials switch Modes switch Navigation keys Back Backspace Enter MODE MAT SCALE OK BS ENTER Hardness scale switch

7.2. Choice of the measurement method.

7.2.1. Rebound method of measuring hardness is appropriate for:

- testing objects weighing more than 5 kg and a wall thickness more than 10 mm;

- massive products, products with a coarse-grained structure, forged and cast products;

- testing objects with minimal preparation of the surface.

7.2.2. UCI method of measuring hardness is appropriate for:

- testing objects with low mass and small wall thickness;

- testing objects with a glossy surface (with special requirements to the minimum size of the imprint);

- surfaces of the testing objects with hardened layers.

8. Operation

8.1. Preparation of the test material.

The surface must be clean and free of oil, grease and dust.

The surface roughness of the material should meet the requirements of a specific probe (paragraph 3.8).

8.2. Features of the methods of measurement of metal hardness:

8.2.1. UCI method:

Distinct reading variations may especially occur with a mass lower than 0.1 kg and a specimen thickness of less than 1 mm if the test material is excited to resonance or sympathetic oscillations.

Otherwise, such test materials must be fixed to a solid base, e.g. using a viscous paste. The same applies to the hardness test blocks.

8.2.2. Rebound method:

With smaller and less heavy test objects, the impact effect of the impact device may cause vibrations which could produce distorted measurement results.

  • Test objects weighing less than 2 kg must in any case be fixed to the support using couplant so that there are no any vibrations.
  • Test objects weighing between 2 kg and 5 kg must be placed on a large metal support (e.g. a table) in such a way that they are not moved or caused to vibrate by the impact.

8.3. Device menu

After connecting the probe, hold TURN ON/OFF button until logo appears on the display:

HARDNESS TESTER

Then, you get to the main menu of the device:

Measuring Calibration Archive Settings Memory card Information

The menu consists of 6 sections:

  1. Measuring
  2. Calibration
  3. Archive
  4. Settings
  5. Memory card
  6. Information

Move through the menu using the navigation keys, for entering

press key . OK

8.3.1. Measuring

Selecting Measuring, you go to the measurement of hardness mode, depending on the probe it will display the angle of the probe at the top of the screen (for rebound probe), or the state of the diamond indenter (for UCI probe).

05:15 cal. off HB Graph Steel DEVICE IS READY FOR OPERATION

Detailed description of the measurement of hardness, see paragraph 9.

8.3.2. Calibration

Selecting Calibration, you go to the table of calibrations where the scale conventionally divided into eight hardness scales: Rockwell (HRC), Brinell (HB), Vickers (HV), Rockwell (HRB), Shore (HS), User 1 (U1), User 2 (U2), User 3 (U3).

Each of the scales can be calibrated to 11 conventional materials: Steel (ST), Alloy Steel (AST), Stainless steel (SST), Cast Iron (CI), Nodular Iron (NI), Aluminums (Al), Brass (BRS), Bronze (BRZ), Copper (CU), User material 1 (U1), User material 2 (U2).

07:31 HRC HB HV HRB HS U1 U2 U3 ST AST SST CI NI AL BRS BRZ CU U1 U2 Empty cell Calibrated User calibration

Detailed description of the calibration process, see paragraph 11.

8.3.3. Archive

Selecting Archive you go to the list of saved measurements, which displays Measurement name, Date and time of the measurement, Scale, Material and Average Value.

PCE Instruments 3500 - Archive - 1

line | Measurement | Value | | ----------- | ----- | | X:59.8 | 59.8 | | R: 4% | 4% | | N:09 | 0.09 |

You can store in memory up to 60 000 measurements.

Scroll through the list using the navigation keys and .

PCE Instruments 3500 - Archive - 2

8.3.4. Settings

Selecting Settings you go to the settings menu for configure the following settings:

Time 13 50 Date 01/01/12 Language ENGLISH Photo OFF Brightness 100% Palette 01 Sound Key+Meas Result Average Volume 50% Auto off OFF Retro OFF Flash OFF Tolerance 7%

Time: time setting.

Date: date setting.

Language: selection language of the device menu (available English and Russian).

Photo: enable / disable the camera (for version with camera).

Brightness: setting the brightness of the display.

Palette: selection of menu themes (creation of color palette of menu makes with special software).

Sound: there are 4 modes of device sound (Off, Key, Measurement, Key and Measurement).

Result: displaying measurement results can be Current (display instantaneous values of measurements) and Average (unit accumulates the series of measurements and display average value of hardness).

Volume: setting the volume of the device.

Auto Off: setting of the automatic switch off device when it is not in use.

Retro: allows you to return to the measurement mode with saved last measurements after restarting of the device.

Flash: enable / disable the cameras flash (for version with camera).

Tolerance, %: This parameter is used only for Smart mode. Set the % value adjusts the range of deviations of measurements to be included in the calculation of the average for the series in Smart mode. Detailed description of the Smart mode, see paragraph 10.4.

Navigate and select options using the navigation keys.

8.3.5. Memory Card

Selecting Memory Card you go to the menu of memory.

8.3.5.1. Create backup copies of calibrations.

11:49 Save Upload Clean CREATE BACKUP COPIES OF CALIBRATIONS 124MB/124MB 0%

After calibration of the probe is recommended to create a backup copy of the calibration (usually the manufacturer makes the calibration of 1-2 scales, to check the probe). This is done in order to be able to resume adequate calibration after incorrect settings in the future.

8.3.5.2. Load backup copies of calibrations.

11:49 Save Upload Clean LOAD BACKUP COPIES OF CALIBRATIONS 124MB/124MB 0%

After the initial save of calibrations you can always download it to the probe. This function is needed for the resumption of adequate calibration in case of wrong settings of the probe.

8.3.5.3. Clear SD card

11 50 Save Upload Clean CLEAR SD CARD 124MB/124MB 0%

Clearing saved records in the archive and backup copies of calibrations: after clearing SD card the archive will be empty and backup copies of calibrations delete. To clean just stored archival records - transmit saved measurements to PC with special software. Thus backup calibrations will be unaffected.

At the bottom of the screen of this menu shown Memory state.

8.3.6. Information

In this menu you can view statistic information.

Measurement Counters: Total: Since last calibration:

Scroll through the list using the navigation keys and .

PCE Instruments 3500 - Information - 2

9. Hardness measurement

9.1. Using of the UCI method

The design of UCI Probe is shown in Figure 1a. Probe has a special removable collapsible nozzle with the puck (Figure 1b). The puck provides the convenience of positioning the probe relative to the testing object and a clip for the measurements

1 2 3 4

Figure 1a. UCI Probe U1 1 – Probe's body; 2 – Collapsible nozzle; 3 – Puck; 4 – Place for fingers

1 4 5 2 3

Figure 1b. Demounted nozzle 1 – Probe's body; 2 – Nozzle main part; 3 – Puck; 4 – Protective tube; 5 – Diamond.

One side of the puck is flat to using probe on flat surfaces. Another side has grooves for using a probe on cylindrical surfaces.

It is marked slots designed for ease of measurement of hardness on cylindrical products of various diameters. Probe with demounted nozzle is usually using to measure hardness in difficult places, such as narrow or deep groove.

The probe may be removed from the nozzle is used for measuring the hardness in the narrow and hard to reach places.

PCE Instruments 3500 - Using of the UCI method - 3

natural_image Close-up of a metallic cylindrical connector with a pointed tip (no visible text or symbols)

Figure 2

Turn on the hardness tester by pressing , and select the measurement mode. For select the measurement mode press

PCE Instruments 3500 - Using of the UCI method - 4

and chose the mode that will be used by keys and

PCE Instruments 3500 - Using of the UCI method - 5

then press . Detailed description of the modes, see graph 10.

PCE Instruments 3500 - Using of the UCI method - 6

flowchart
graph TD
    A["Presence of user calibration"] --> B(cal. off)
    C["The state of the diamond indenter (ready for measurement)"] --> B
    D["Mode"] --> E["Graph"]
    E --> F["05:15 HB"]
    F --> G["Scale"]
    F --> H["Steel"]
    I["DEVICE IS READY FOR OPERATION"] --> J["i"]
    J --> K["Material"]

Then select the scale and material hardness for which there is a proper calibration (how to calibrate device see paragraph 11).

For select the hardness scale press and chose the scale that will be used by keys and then press.

cal. off Graph HB HRC HB HV HRB HS DEVICE IS READY FOR OPERATION

For select the material press and chose the material that will be used by keys and then press.

PCE Instruments 3500 - Using of the UCI method - 8

line | Grade | Value | |-------|-------| | 1 | 47.7 | | 2 | 47.7 | | 3 | 47.7 | | 4 | 47.7 | | 5 | 47.7 | | 6 | 47.7 | | 7 | 47.7 | | 8 | 47.7 |

During the measurement you can select other hardness scale. The displayed measurement reading will be converted according with the new hardness scale if it is calibrated.

NOTE! Calibration is carried out by the direct method, so the conversion is carried out on the basis of pre-calibration, and does not correspond to any standard.

Install the probe puck on the sample surface, keeping it in the tough skirt as shown in Figure 3a. By clicking on the skirt thrust both hands to bring the diamond tip of the probe perpendicular to the sample surface to the touch (Figure 3b). Slowly (in about 0.5 seconds) by pressing with a force of (5 or 1 kg depending on probe type) thrust skirt, push the diamond tip into the metal surface, preventing swinging (Figure 3c). After the sound signal, remove the probe from the tested object.

90° a b c

Figure 3

NOTE! For correct operation of the UCI probe try not to exceed the pressing force which is correspond to the probe (1 or 5 kg).

The display shows the value of hardness. The result of measurement is displayed on the display until the next measurement.

cal. 1p HRC 48.6 Graph Steel This symbol indicates that the diamond indenter is in contact with the testing object or device is not ready for the next measurement X:47.7 R: 4% N:02 45.5

WARNING! Not allowed a sharp click and scratching sample by indenter, this may lead to exceeding the allowable value of error and damage the diamond indenter.

NOTE! The "exciter" in the probe is powered from the batteries. If the batteries are depleted you will get erratic test results.

You can get the value of the current measurement or the average for the series of measurements, depending on the settings in Settings menu Current or Average:

11:01 Time 11:01 Date 02/01/12 Language ENGLISH Photo OFF Brightness 50% Palette 01 Sound Key+Meas Result Current Volume Current Auto off Average Retro OFF Flash OFF Tolerance 7%

After the measurement, you can save measurement (series of measurements) in the archive by pressing ENTER key .

PCE Instruments 3500 - Using of the UCI method - 12

line | Time | Value | |------|-------| | 2 | 95.1 | | 8 | 95.1 |

Keyboard appears:

cal. off HB 95 Graph Steel MEASUREMENT 1 2 3 4 5 6 7 8 9 0 < Q W E R T Y U I O P " A S D F G H J K L _ ? Z X C V B N M @ # $ / 85 0 2 4 6 8

Move the cursor over the keyboard with navigation keys and select

symbols by pressing . Then press for saving. The record is stored in the archive.

9.2. Using of the Rebound (Leeb) method

Select scale and material with the existing calibration.

PCE Instruments 3500 - Using of the Rebound (Leeb) method - 1

PCE Instruments 3500 - Using of the Rebound (Leeb) method - 2

Press the buttons and to set the angle of the measurement. Fixed angle is displayed in the upper part of the display.

Angle of the measurements cal. 1p HB Graph Steel DEVICE IS READY FOR OPERATION

Position of the probe perpendicular to the ground plane corresponds to the angle 0^ .

Place the rebound probe vertically onto the test surface and press it slightly against the surface. Charge the impact body by a charging mechanism pressing the probe enclosure until the clicks, as shown below:

PCE Instruments 3500 - Using of the Rebound (Leeb) method - 4

natural_image Two identical cylindrical mechanical components with black bases and threaded shafts, one showing a downward arrow (no text or symbols)

Figure 4. Rebound probe

Press the release button on top of the impact device by finger of your free hand.

After pressing the shutter release button and the impact body hitting in the area of measurement sounds a beep and the display shows the value of measured hardness.

PCE Instruments 3500 - Using of the Rebound (Leeb) method - 5

line | X | Y | | ---- | ---- | | 384.3| 384.3|

WARNING! The minimum distance between the imprints of measurements must be not less than 3 mm.

10. Measurement modes

To selecting the measurement mode go to the Measuring and press

PCE Instruments 3500 - Measurement modes - 1

The device will offer you the following measurement modes:

  • Graph – mode of the construction graph;
  • Histogram – mode of the construction columnar bar chart;
  • Statistic – statistics mode;
  • Smart – mode of filter incorrect measurements;
  • Signal – the display signal mode (only for Leeb).

cal. 2p 09:16 HRC Histogram Graph Histogram Statistic Smart Steel i IS READY FOR OPERATION

To select the measurement mode, press

PCE Instruments 3500 - Measurement modes - 3

10.1. Graph mode

PCE Instruments 3500 - Graph mode - 1

line | X | R | N | | ---- | ---- | --- | | 29.9 | 0% | 01 | | 29.9 | | | | | | | | | | 29.9 |

In graph mode, the device displays the current measurement value or average of series of measurements, depending on selected the Current or Average in Measurement settings menu:

11:01 Time 11:01 Date 02/01/12 Language ENGLISH Photo OFF Brightness 50% Palette 01 Sound Key+Meas Result Current Volume Current Auto off Average Retro OFF Flash OFF Tolerance 7%

10.2. Histogram mode

The device builds histogram of series measurements.

PCE Instruments 3500 - Histogram mode - 1

10.3. Statistics mode

Statistics mode allows to monitoring the following parameters of measurement series: Maximum, Minimum, Deviation, Average, Number of measurements.

cal. 1p HRC 62.2 Statistic Steel Maximum 62.2 Minimum 61.7 Deviation 0.3 Average 62.1 Var. coeff. 0.8 Num of meas. 5

10.4. Smart mode

Smart mode allows the user to identify the general sequence of measurements. The device selects the first three series of measurements that do not exceed the specified tolerance. After that, the following measurements that exceed a given tolerance will be excluded from the series and will not be taken into account when calculating the average of the series.

cal. off HB 113 Smart Steel 115 110 105 100 95 0 2 4 6 8

Filled with the color readings indicate that the device has fixed the general sequence and Smart mode is active.

PCE Instruments 3500 - Smart mode - 2

line | Time | Value | |------|-------| | 0 | 107 | | 2 | 112 | | 4 | 104 |

To set the tolerance of Smart Mode, go to Settings:

And set the percentage tolerance values from 1 to 10.

12 05 Time 12 05 Date 02/01/12 Language ENGLISH Photo OFF Brightness 60% Palette 01 Sound Key+Meas Result Current Volume 50% Auto off OFF Retro OFF Flash OFF Tolerance 7% 8% -

10.5. Signal mode

Mode is active only for Leeb probe and shows the voltage produced during the impact and rebound of impact body. M is the maximum value of the conditional code number which corresponds to the received signal.

PCE Instruments 3500 - Signal mode - 1

line | Time | Signal | |------|--------| | 0 | -10 | | 3 | 5 | | 6 | 28 | | 9 | -25 | | 12 | -10 |

11. Calibration

11.1. Main calibration

To calibrate the probe you will need 3 samples of material with a known hardness. Hardness range should be wider than the hardness of the materials which will be measured (The values should be the maximum or more, minimum or less and average).

Select in the main menu Calibration:

The device goes to the table, each cell of the table corresponds to the certain scale of calibration for the certain material:

Rockwell C Brinell Vickers Rockwell B Shore User 1 User 2 User 3 Steel Alloy Steel Stainless Steel Cast Iron Nodular Iron Aluminums Brass Bronze Copper User material 1 User material 2 HRC HB HV HRB HS U1 U2 U3 ST AST SST CI NI AL BRS BRZ CU U1 U2 Empty cell Calibrated User calibration

Each of the scales can be calibrated to 11 conventional materials: Steel (ST), Alloy Steel (AST), Stainless steel (SST), Cast Iron (CI), Nodular Iron (NI), Aluminum (Al), Brass (BRS), Bronze (BRZ), Copper (CU), User material 1 (U1), User material 2 (U2).

All calibrations of the device can be calibrated for any materials and any scales, and are divided in this way just for practicality.

Press the navigation keys to choice cell for calibration, for example HRC for Steel:

HRC HE HV HRB HS U1 U2 U3 ST AST BST CI NI AL BRS BRZ CU U1 U2 Empty cell Calibrated User calibration 1431

For select press , the table appears:

R1: 200.0 C1:200 HRC R2: 500.0 C2:500 ST R3: 700.0 C3:700 X: 000 60 50 39 29 19 1000 1340 1680 2020 2360

The device, making the measurement, gets the nominal codes, the purpose of calibration - is to find the correlation between the nominal code and hardness values (construction of the relation function).

To start the calibration, enter the real values of the samples by pressing . ENTER

PCE Instruments 3500 - Main calibration - 4

for setting real values of Use the navigation keys hardness, to go to the next digit number, press

xt digit number, press :

PCE Instruments 3500 - Main calibration - 6

line | Time | Value | |------|-------| | 1000 | 19 | | 1340 | 39 | | 1680 | 60 | | 2020 | 39 | | 2360 | 60 |

For ending setting of first value press again , then the button

PCE Instruments 3500 - Main calibration - 7

for adjusting the values of the next samples.

Then, move your selection on the hardness value according to the sample, and make at least 5 measurements. Make sure that the value of X (current value of the code) would not varied by more than 3% .

PCE Instruments 3500 - Main calibration - 8

line | Time | Value | |-------|-------| | 1000 | 60 | | 1340 | 50 | | 1680 | 29 | | 2020 | 19 | | 2360 | 19 |

If you get an obvious error measurement, press and the last measurement will be removed from the series.

Go to the next nominal pressing, and make the same procedure with other samples, you end up with:

PCE Instruments 3500 - Main calibration - 9

line | x | y | | ---- | ---- | | 1000 | 60 | | 1340 | 39 | | 1680 | 29 | | 2360 | 19 |

To check the correctness of the got values, move the cursor to the X:000 and take a few measurements on one of the sample.

For save, press , window will appear:

SAVE? BACK No Yes

Press YES by . Calibration is saved. Select the appropriate material and scale, and device is ready for operation.

11.2. User (additional) calibration

Each of the saved main calibration may be further corrected. Additional calibration is recommended in the following cases:

  • If the measurements of the device on the samples are constant, but differ from the nominal value of the standard block;
  • After extended storage (more than 3 months.);
  • After intensive operation;
  • With a significant change in the operating conditions (temperature, humidity, etc.).

For the calibration of hardness scale needed one (single-point calibration) or two (two-point calibration) standard hardness tests samples with the maximum and minimum values for the controlled range of the hardness.

For example, we have two steel sample of known hardness of HRC, and device shows a stable deviation for hardness measurements on it. For making user two-point calibration select in main menu

Calibration:

ST AST SST CI NI AL BRS BRZ CU U1 U2 HRC HB HV HRB HS U1 U2 U3 Empty cell Calibrated User calibration 14:55

OK Press and the screen appears:

PCE Instruments 3500 - Calibration: - 2

Press ENTER for select the number of calibration points, as we have

two sample - set 2 with pressing buttons and, then again ENTER

PCE Instruments 3500 - Calibration: - 3

line | X | Y | |-------|----| | 1000 | 60 | | 1340 | 50 | | 1680 | 39 | | 2020 | 29 | | 2360 | 19 |

Move the cursor by pressing to select the first row. Take about 5 measurements on the first sample, the instrument will display the average of the series in accordance with the current calibration. If

you get an obvious error measurement, press and the last measurement will be removed from the series. You will get:

PCE Instruments 3500 - Calibration: - 4

line | X | Y | |-------|-----| | 1000 | 60 | | 1340 | 50 | | 1680 | 39 | | 2020 | 29 | | 2360 | 19 |

Press and by pressing and set the nominal value

of the sample. To go to the next digit number, press , for saving first value press . After adjusting the value of the first sample you get:

PCE Instruments 3500 - Calibration: - 5

line | Time | Value | |-------|-------| | 1000 | 60 | | 1340 | 50 | | 1680 | 39 | | 2020 | 29 | | 2360 | 19 |

PCE Instruments 3500 - Calibration: - 6

Press to adjust the second sample value, and perform the same operation. In the end, get:

PCE Instruments 3500 - Calibration: - 7

line | X | Y | |-------|------| | 1000 | 60 | | 1340 | 50 | | 1680 | 29 | | 2020 | 19 | | 2360 | 19 |

To save, press , window will appear:

SAVE? BACK No Yes

Press YES by . Calibration is saved. Select the appropriate material and scale, and device is ready for operation.

About the stored user calibration will symbolize the next state of the cell:

16 39 HRC HE HV HRB HS U1 U2 U3 ST AST SST GCI NCI AL BRS BRZ CU U1 U2 Empty cell Calibrated User calibration

To delete user calibration, go to the user calibration mode and set to 0 for N:

PCE Instruments 3500 - Calibration: - 10

12. Technical maintenance, special conditions of operation

12.1 On the whole, hardness testers do not require any special maintenance. However, for the purpose of hardness tester stable operation, regular maintenance is advisable.

12.2. Probe maintenance

Clean hardmetal ball and diamond pyramid from dust, mud and oil traces. Use soft cloth impregnated with alcohol solution.

Check the probe operation regularly by conducting hardness measurements on hardness reference blocks. Do not use reference test block with expired period between verifications (more than 2 years).

12.3. Information processing unit maintenance

To clean from any pollution, use soft dry cloth. Do not use the water, since the hardness tester is neither spray-proof nor water-proof due to the joints on its body.

Do not use any solvents, they can damage indication signs and writings on the front and back sides of the body.

12.4. Battery maintenance

The battery average life is not less than 3 years. The battery used in compliance with the "C" or "AA" international standard. It is done for the convenience of its replacement when it is required or sharp reduction of the continuous operation time (paragraph 3.6) independently of the country. Replacement is possible only by the battery with similar characteristics in compliance with the marking on it. From environmental protection point of view, the best thing is to use the battery.

12.5. Storage

12.5.1. Hardness tester shall be kept in the carry case, the probe and the battery shall be disconnected.

12.5.2. If hardness tester is kept in the carry case for than 14 days, the battery shall be taken out from its compartment in the information processing unit.

12.5.3. It is recommended to keep hardness testers in closed premises with the relative humidity not more than 80%, there shall be no mold, paints, acids, chemical agents and other chemicals, the evaporation of which my give a harmful effect. Sharp fluctuations of temperature and humidity which can result in dew formation are not allowed.

12.6. Transportation

12.6.1. Hardness tester transportation in the carry case shall be only in closed vehicles, where the possibility of mechanical damage or atmospheric precipitation is excluded.

12.6.2. The way packed in carry cases hardness testers are located inside the vehicle shall exclude.

12.7. Putting into operation after storage and transportation

12.7.1. After storage or transportation under the temperature lower than -5^ , before starting hardness tester operation, it is necessary to keep it not longer than 1 hour under the temperature higher than +10^ and not less than 2 hours under the temperature higher than 0^ .

12.7.2. Before operating hardness tester which was stored for more than 3 months and transported for more than 2 months, it is necessary to check such hardness tester on the reference hardness test blocks. If the measured AVEARGE value of the hardness tester does not correspond to the reference hardness test block NOMINAL value within the error limits, it is necessary to calibrate the hardness tester.

12.8. Special operation conditions

12.8.1. Increased dust content and humidity. Put the information processing unit of hardness tester into a transparent plastic bag. Tighten it at the level of connective cable a bit lower than the probe plug.

After the work under such conditions is finished, information processing unit shall taken out of the plastic bag and air it.

12.8.2. Frost (<0°C). Information processing unit is the most sensitive to low temperature part of hardness testers, especially LCD. If there is a possibility, keep hardness tester closer to your body and protect id with your coat or keep in the inside pocket, taking it out time from time for inputting the data into the archive.

13. Precautions and trouble shouting

13.1. Treat the hardness tester with care. Any wrong treatment may result in the violation of the present Technical Reference and Operation Manual regulations and, thus, lead to the manufacturer hardness tester warranty cessation.

13.2. Always check the integrity of the cables, Information processing unit and probes. Provide immediate replacement of the damaged parts by the original ones. This job shall be performed by skilled personnel.

13.3. Do not expose the hardness tester to aggressive chemical medium.

13.4. Do not leave the hardness tester in the direct Sun.

13.5. Do not sink the hardness tester into any liquids. If the hardness tester gets wet, take the battery out and leave for 24 hours to get dry. If the hardness tester is used under the increased humidity or dust conditions, place the information processing unit into the plastic bag. After work period is over, it is mandatory to get the hardness tester dry.

14. Manufacturer's guarantee and service maintenance

14.1. In case when hardness tester (information processing unit and probes), supplied by the Seller within one year since the delivery date used properly, appear defective, the Seller shall repair or replace them or supply new parts therefore and send them to the Buyer. The other parts of the device (battery, charger, bag) warranty does not cover. The Buyer must prepay any shipping charges, taxes, or duties associated with transportation of the product to the service location, and Seller pays return shipping and associated costs. In addition, the Buyer shall be responsible for insuring any product shipped or returned to an authorized service location, and the Buyer shall bear risk of loss during shipping to the service location, and Seller all risks during delivery to the Buyer. In order to exercise its rights the Buyer shall inform the Seller as soon as possible after the date when such defect appeared.

14.2. Prepare technically grounded reclamation document and send it together with the hardness tester to the guarantee service or to the manufacturer at the address:

14.3. Post guarantee maintenance is conducted by the manufacturer service centre upon the Customer request.